ELECTRO SURGICAL UNITS – Meqube https://www.meqube.com Refurbished Medical Equipment Sales and Service Fri, 21 Jan 2022 04:40:55 +0000 en-US hourly 1 https://wordpress.org/?v=6.6.2 https://www.meqube.com/wp-content/uploads/2022/01/cropped-logo-32x32.jpg ELECTRO SURGICAL UNITS – Meqube https://www.meqube.com 32 32 MEDTRONIC FORCE TRIAD https://www.meqube.com/product/medtronic-force-triad/ Wed, 19 Jan 2022 13:23:12 +0000 https://www.meqube.com/?post_type=product&p=526 Features The Covidien Force Triad energy platform is a full-featured electrosurgical system that provides electrosurgical cutting and coagulation, bipolar functionality, and vessel sealing in a single generator. This Electrosurgical Unit is intended for open and laparoscopic procedures and includes; TissueFect, ValleyLab, and LigaSure fusion technologies. It features a user-friendly interface with three touchscreens that allow the user to control the system functions.
  • TissueFect sensing technology to precisely manage energy delivery.
  • Enhanced LigaSure tissue fusion technology that fuses vessels up to 7mm, lymphatics, pulmonary vasculature and tissue bundles.
  • alleyLab mode that provides a unique combination of monopolar hemostatis and dissection.
  • Automatic instrument recognition.
  • Compatible with conventional electrosurgical instruments and all current and new LigaSure instruments.
]]>

Specifications

General
  • Output Configuration: Isolation output.
  • Cooling: Natural convection and fan.
  • Display: Three LCD touchscreens.
  • Connector ports: LED illuminated Smart connector readers
    • ForceTriad energy platform cart (FT900) Universal.
    • Mounting cart (UC8009), and/or the UC8010 Over-shelf.
    • Operating room boom systems.
    • Any Stable, flat surface such as a table or cart top.
Equipment Dimensions and Weight
  • Height: 25.5 cm (10 in.)
  • Width: 45.8 cm (18 in.)
  • Depth: 50.8 cm (20 in.)
  • Weight: 13.6 kg (30 lbs)
Activation Tone
  • The audio levels stated below are for activation tones (cut, Valleylab, coag, bipolar, and LigaSure Modes) and alarm tones (REM and system alarms) at a distance of one meter.
Volume and Frequency
  • Volume (adjustable): 45 to 65 dBA
  • Frequency Cut: -660 Hz
    • Valleylab: 800 Hz.
    • Coag: 940 Hz.
    • Bipolar: 940 Hz.
    • LigaSure: 440 Hz.
  • Duration: Continuous while the system is activated.
Alarm Tone
  • Volume (adjustable): >65 dBA.
  • Frequency: -REM: 660 Hz.
    • Regrassp: Two tones: High= 985 Hz, Low = 780 Hz.
    • Seal Complete: 985 Hz.
    • Error/System Alert: Beep tone = 1400 Hz.
    • Duration:
      • REM: Two 1/2 second tones separated by 1/2 second for each REM event
      • Reactivate/Regrassp: Four 175 msec tones-high, low, high–low separated by 1/2 second.
      • Check Instrument: Six 175 msec tones – high, low, high low, high, low -Seal Complete: Two 175 msec tones separated by 175 msec for each Seal Complete event.
      • Error/System Alert: Two 1/2 second tones separated by 1/2 second for each Error/ System Alert event.
REM Contact Quality Monitor
  • Interrogation frequency: 80 kHz ± 10 kHz
  • Interrogation current: < 100μA
  • Interrogation voltage: < 12V RMS
Acceptable Resistance Range
  • REM resistance measurements are ± 10% during RF activation and ± 5% when RF output is not activated.
  • REM patient return electrode: 5 to 135 ohms or up to a 40% increase in the initial measured contact resistance (which ever is less).
  • If the measured resistance is outside the acceptable range(s) noted above, a REM fault condition occurs.
REM Alarm Activation
  • REM patient return electrode: When the measured resistance exceeds the standard range of safe resistance (below 5 ohms or above 135 ohms) or when the initial measured contact resistance increases by 40% (whichever is less), the REM alarm indicator enlarges and flashes red and yellow, a tone sounds twice, and RF output is disabled. The indicator remains illuminated red and yellow until you correct the condition causing the alarm. Then, the indicator illuminates green and RF output is enabled.
Autobipolar
  • The ForceTriad energy platform is equipped with an autobipolar feature that allows for automatic activation of bipolar energy.
  • Note: The autobipolar electrode function requires the use of the Valleylab E0018 bipolar Instrument cord.
  • Autobipolar Specificaitons:
    • Interrogation frequency: 80 kHz ± 10 kHz.
    • Interrogation current: < 100 μA.
    • Activation impedance: 20 Ω to 500.
    • Deactivation impedance: User selectable: 1,500 Ω 1,800 Ω or 2,200 Ω.
    • Measurement accuracy: ± 14% of Full Scale activation impedance while keying active above 10 watts.
    • + 50%, -14% of full scale activation impedance while keying active at 10 watts or below.
    • ± 5% of Full Scale activation impedance while keying inactive.
    • Keying delay: User selectable in 500 msec increments from 0 sec to 2.5 sec
Duty Cycle
  • Under maximum power settings and rated load conditions, the ForceTriad energy platform is capable of operating a duty cycle of 25% defined as 10 seconds active and 30 seconds inactive, in any mode for a period of 4 hours.
  • Caution
    • Use of duty cycles greater than 25% (10 seconds active followed by 30 seconds inactive) will increase the risk that heat build-up under a return electrode may be high enough toinjure the patient. Do not continuously activate for longer than one minute.
Low Frequency (50/60 Hz) Leakage Current
  • Enclosure source current, ground open <300 μA.
  • Source current, patient leads, all outputs
    • Normal polarity, intact ground: < 10 μA.
    • Normal polarity, ground open: < 50 μA.
    • Reverse polarity, ground open: < 50 μA.
    • Mains voltage on applied part: < 50 μA.
  • Sink current at high line, all inputs<50 μA
Input Power

100-120 Volt

  • Maximum VA at nominal line voltage:
    • Idle: 52 VA
    • Bipolar: 450 VA
    • Cut: 924 VA
    • Coag: 530 VA
  • Input Mains Voltage, Full Registration Range:
    • 90-132 Vac
  • Input Mains Voltage, Operating Range:
    • 85-132 Vac

220-240 Volt

  • Maximum VA at nominal line voltage:
    • Idle: 52 VA
    • Bipolar: 450 VA
    • Cut: 924 VA
    • Coag: 530 VA
  • Input Mains Voltage, Full Registration Range:
    • 208-264 Vac
  • Input Mains Voltage, Operating Range:
    • 170-264 Vac

100-120 Volt

  • Mains Current (Maximum):
    • Idle: 0.4 A
    • Bipolar: 2.0 A
    • Cut: 7.0 A
    • Coag: 4.0 VA
    • LigaSure: 5.0 VA
  • Mains Line Frequency Range (Nominal):
    • 50 Hz to 60 Hz
  • Fuses (2): 5mm x 20 mm A8, 250 V:
    • Fast Blow
  • Power cord: 3-prong hospital grade:
    • Connector

220-240 Volt

  • Mains Current (Maximum):
    • Idle: 0.2 A
    • Bipolar: 1.0 A
    • Cut: 3.5 A
    • Coag: 2.0 VA
    • LigaSure: 2.5 VA
  • Mains Line Frequency Range (Nominal):
    • 50 Hz to 60 Hz
  • Fuses (2): 5mm x 20mm A8, 250 V:
    • Fast Blow
  • Power cord: 3-prong hospital grade:
    • Connector
]]>
COVIDIEN VALLEY LAB FORCE FXc https://www.meqube.com/product/covidien-valley-lab-force-fxc/ Wed, 19 Jan 2022 12:18:09 +0000 https://www.meqube.com/?post_type=product&p=521 Features The Covidien Force FX-C is an electrosurgical generator, also referred to as the "Covidien Force FX electrosurgical generator C " , that provides isolated output power for cutting, desiccating, and fulgurating tissue during bipolar and monopolar surgery. This Valley lab ESU automatically senses resistance and adjusts the output voltage to maintain a constant effect across different tissue densities. This adjustment is only available in bipolar or cut modes and is based on the power setting and the level of tissue resistance. It features 3 bipolar modes, 3 monopolar cut modes, 3 monopolar coag modes, and supports simultaneous coagulation. The Covidien Force ESUs support the use of CUSA System 200 CUSA EXcel+ , and CUSA handpieces with the CUSA electrosurgical module.
  • Instant Response technology ensures that the power delivered remains virtually constant, regardless of the tissue type
  • Improved performance at lower power settings minimizes the risk of tissue damage and neuromuscular stimulation
  • Three internal microcontrollers reduce system reaction time and increase the system's processing speed
  • Spray coagulation voltage of no more than 9000 volts peak-to-peak output for broad, but superficial coagulation with limited capacitive coupling
  • A Power Efficiency Rating (PER) of approximately 98 for accurate and consistent cut performance
  • Three cut modes, all controlled by Instant Response™ technology, offer surgeons a variety of choices:
    • Low Cut for delicate tissue or laparoscopic cases
    • Pure Cut for a clean, precise cut
    • Blend for cutting with hemostasis
  • Four coag modes:
    • Desiccate for low voltage contact coagulation suitable in laparoscopic and delicate tissue work
    • Fulgurate (high crest factor) for efficient noncontact coagulation in most applications
    • Fulgurate (low crest factor) for lower voltage coagulation requirements
    • Spray for coagulating large tissue areas with superficial depth of necrosis
  • Three bipolar modes:
    • Precise, Standard, and Macrobipolar are controlled by the Instant Response™ system
    • Precise and Standard setting utilize low voltage to prevent sparking
  • Versatile system that is uniquely compatible with other devices, including:
    • Force Argon™ II and Force GSU™ argon coagulation system
    • CUSA EXcel™ and CUSA™ 200 ultrasonic surgical aspirators
    • OptiMumm™ smoke evacuator, through a direct cable link
    • Valleylab VLCM bipolar current monitor
  • Compatible with and used as the electrosurgical energy source for:
    • Dyonics* Control RF arthroscopic ablation system
    • Dyonics* Electroblade™ rotary resection system
    • Cook Vascular Perfecta™ EDS pacemaker lead extraction system
  • Compatible with, and the exclusive electrosurgical generator for, the Computer Motion Hermes™ Voice Command system
]]>
Features The Covidien Force FX-C is an electrosurgical generator, also referred to as the "Covidien Force FX electrosurgical generator C " , that provides isolated output power for cutting, desiccating, and fulgurating tissue during bipolar and monopolar surgery. This Valley lab ESU automatically senses resistance and adjusts the output voltage to maintain a constant effect across different tissue densities. This adjustment is only available in bipolar or cut modes and is based on the power setting and the level of tissue resistance. It features 3 bipolar modes, 3 monopolar cut modes, 3 monopolar coag modes, and supports simultaneous coagulation. The Covidien Force ESUs support the use of CUSA System 200 CUSA EXcel+ , and CUSA handpieces with the CUSA electrosurgical module.
  • Instant Response technology ensures that the power delivered remains virtually constant, regardless of the tissue type
  • Improved performance at lower power settings minimizes the risk of tissue damage and neuromuscular stimulation
  • Three internal microcontrollers reduce system reaction time and increase the system's processing speed
  • Spray coagulation voltage of no more than 9000 volts peak-to-peak output for broad, but superficial coagulation with limited capacitive coupling
  • A Power Efficiency Rating (PER) of approximately 98 for accurate and consistent cut performance
  • Three cut modes, all controlled by Instant Response™ technology, offer surgeons a variety of choices:
    • Low Cut for delicate tissue or laparoscopic cases
    • Pure Cut for a clean, precise cut
    • Blend for cutting with hemostasis
  • Four coag modes:
    • Desiccate for low voltage contact coagulation suitable in laparoscopic and delicate tissue work
    • Fulgurate (high crest factor) for efficient noncontact coagulation in most applications
    • Fulgurate (low crest factor) for lower voltage coagulation requirements
    • Spray for coagulating large tissue areas with superficial depth of necrosis
  • Three bipolar modes:
    • Precise, Standard, and Macrobipolar are controlled by the Instant Response™ system
    • Precise and Standard setting utilize low voltage to prevent sparking
  • Versatile system that is uniquely compatible with other devices, including:
    • Force Argon™ II and Force GSU™ argon coagulation system
    • CUSA EXcel™ and CUSA™ 200 ultrasonic surgical aspirators
    • OptiMumm™ smoke evacuator, through a direct cable link
    • Valleylab VLCM bipolar current monitor
  • Compatible with and used as the electrosurgical energy source for:
    • Dyonics* Control RF arthroscopic ablation system
    • Dyonics* Electroblade™ rotary resection system
    • Cook Vascular Perfecta™ EDS pacemaker lead extraction system
  • Compatible with, and the exclusive electrosurgical generator for, the Computer Motion Hermes™ Voice Command system
]]>