CAREFUSION LTV 1200

The Vyaire LTV 1200 is a portable ventilator. LTV Ventilator can be used in-home, intra-hospital transport, or in emergency infield use. Vyaire System comes with preset buttons for infants, pediatrics, and adult patients and can be customized per each patient’s needs. It offers both invasive and non-invasive modes of ventilation in Pressure Control, Pressure Support, Volume Control, and Spontaneous Breath Types. With the Vyaire Turbine technology, the LTV 1200 eliminates the need for a high-pressure air source.

  • Quick Patient Ventilation – Choose infant, pediatric, or adult to quickly configure initial ventilation.
  • More versatile than transport ventilators and significantly smaller than ICU systems.
  • The O2 Conserve feature reduces O2 consumption when the time and place of care require it.
  • The internal battery allows transfer from one power sources to the next without interruption.
Share:

Specifications

Dimensions
  • Height: 3.25 In (8.4 cm)
  • Width: 10.5 In (27 cm)
  • Depth: 13.5 In (38 cm)
  • Weight: 14.5 lbs (6.5 kg)
Power
  • Ventilator input 11–15 VDC
  • AC adapter input 100–250 VAC
  • 50–60 Hz
Pneumatic
  • High pressure O2 source 40–80 PSIG (2.8–5.5 bar)
  • Low pressure O2 source < 80 LPM, < 10 PSIG
Controls
  • Ventilation Modes: Control, Assist/Control, SIMV/CPAP and NPPV
  • Breath Types: VC, PC, PS and spontaneous
  • Breath Rate: 0–80 BPM
  • Tidal Volume: 50–2,000 mL
  • Inspiratory Time: 0.3–9.9 sec
  • PC/PS/Spontaneous Flow: 160 LPM
  • Pressure Control: 1–99 cmH2O
  • Pressure Support: 1–60 cmH2O
  • Sensitivity: Off, 1–9 LPM
  • O2%: 21–100%
  • O2 flush: 1–3 min
  • Low pressure O2: On/Off
  • PEEP/CPAP: 0–20 cmH2O
  • Insp/Exp hold: 6 seconds maximum
  • Manual breath: 1x current settings
Monitors and Controls
  • Peak Inspiratory Pressure: 0–120 cmH2O
  • Mean Airway Pressure: 0–99 cmH2O
  • PEEP: 0–99 cmH2O
  • Breath Rate: 0–250 BPM
  • Airway Pressure Display: -10–108 cmH2O
  • Exhaled Tidal Volume: 0–4,000 mL
  • Exhaled Minute Volume: 0–99.9 L
  • I:E Ratio: 99:1–1:99
  • Calculated Peak Flow: 10–100 LPM
  • AutoPEEP: 0–100 cmH2O
  • Static Compliance: 1–999 mL/cmH2O
  • Patient Effort: Green LED
Extended Features
  • SBT
  • Ventilator presets (infant, pediatric and adult)
  • Variable rise time
  • Variable flow termination
  • Variable time termination
  • Pressure control flow termination
  • Leak compensation
  • O2 conserve
  • O2 cylinder duration calculation
  • O2 flush
  • Apnea interval
Fixed Alarms
  • Disconnect/Sense
  • External power low and lost
  • High and low O2 inlet pressure
  • Internal battery low and empty
  • Ventilator inoperative