MAQUET DATASCOPE CS300

Features

The Maquet Datascope CS300 IABP with IntelliSense combines fiber-optic speed with automatic in vivo calibration. The result is a faster time for therapy, faster signal acquisition, and faster adaptation to rate and rhythm changes. Maquet has the only fiber-optic IABP and catheter system that automatically calibrates in the patient after insertion and automatically recalibrates in vivo every two hours or sooner should the patient or environmental conditions change.

The CS300 Datascope IABP with INTELLISENSE™ software is designed to emulate an experienced user, allowing you more time to focus on your patient. This sophisticated algorithm automatically selects the best lead and trigger source, sets optimal inflation and deflation timing, and still allows the clinician the ability to fine-tune deflation timing. The Maquet IABP quickly adapts to rate or rhythm changes in order to optimize augmentation and support during diastole.

 

  • Faster inflation and deflation speed
  • CS300 has true one-button start-up
  • Automatically calibrates the fiber-optic pressure sensor in the patient and recalibrates every 2 hours or sooner should patient or environmental conditions change
  • Automatically evaluates and selects the best lead and trigger source
  • Automatically sets optimal timing
  • Automatically adjusts to changes in patient conditions without clinician intervention
  • A Maquet fiber-optic IAB catheter
  • Conventional fluid-filled IAB catheters
Share:

Specifications

Dimensions
  • Size on Cart: 43.1” H x 22.3” D x 16.8” W (109 cm H x 56.6 cm D x 42.7 cm W)
  • Size off Cart: 26.9” H x 20.5” D x 10.8” W (68.3 cm H x 52.1 cm D x 27.4 cm W)
  • Console Weight: 84.8 lbs. (38.4 Kg) nominal (includes S.D., CRM, and He tank)
  • Monitor: 9.5 lbs. (4.3 Kg) nominal
  • Hospital Cart Weight: 52.4 lbs. (23.8 Kg) nominal
  • Internal Battery: 34 lbs. (15.4 Kg) nominal
  • Storage Bag: 4.6 lbs. (2.1 Kg) nominal (includes doppler and doppler holder))
Color Display
  • Color TFT Liquid Crystal Display (LCD) 8.3”(21cm)W x 6.2”(15.8cm)H; up 45°, 10.4” diagonal (26.4 cm) down 55°, right 70°, left 70° viewing angle; Rotates 330°; Tilts 180°; Detachable; Laptop-like closure for storage and protection; Remote monitor mount (optional)
Trigger
  • ECG Trigger: Threshold dynamically adjusted by the system for high sensitivity and selectivity of the R-wave detection; Minimum = 120µV ± 20µV at a normal gain; 40µV at max. gain
  • Pressure Trigger: Default trigger threshold is automatically adjusted to 38% of the systolic pulse height; 7 mm minimum
  • Manual threshold mode: User adjustable between 7 and 30 mmHg ± 3 mmHg
  • Pacer A Trigger: R-wave detection (as above) except pacer blanking is extended to 100 ms
  • Pacer V/A-V Trigger: V Pacer: fixed at a rate up to 185 bpm (no demand pacing)
  • A-V Pacer: fixed at a rate up to 125 bpm (no demand pacing) with A-V intervals between 80-224 ms
Power
  • Mains Voltage: 100-120 VAC ± 10% or 220-240 VAC ± 10%
  • Mains Frequency: 50/60 Hz ± 3 Hz
  • Internal Battery: 24 VDC (nominal), 17.2 Amp-hour, approx. 3 hrs. @ 90 bpm
  • Battery Type: Maintenance free; Sealed lead-acid
Pressure
  • Pressure Output: (electrically isolated)
  • Pressure Range: 0 to +300 mmHg (minimum)
Operation
  • Op. Temp.: 10°C – 40°C
  • Op. Humidity: 5 – 95% (R.H.) non-condensing
  • Op. Altitude: 0 – 12,000 feet (3,657 m); automatic altitude correction for IAB pressure
ECG
  • ECG Leads: In Auto Operation Mode: I, II, III, External In Semi-Auto Mode: I, II, III, AVR, AVL, AVF, V, External (12 lead compatibility)
  • ECG Gain (default): 1 V output per 1 mV input ±5% (waveform automatically scaled to occupy ECG display window)
  • Gain (variable): 0.15 to 3.0 cm/mV ±20% (autoscaling disabled)
  • Frequency Response: 0.5-12 Hz (display); 0.5-135 Hz (Output to External Monitor)
  • Defibrillator Protection: Discharge level ≤ 360 J (trace returns to screen in 5 sec max)
  • ESIS: Automatic suppression with internal ECG amplifier