PHILIPS TRILOGY 200

Features

The Philips Respironics Trilogy 200 portable ventilator is lightweight and easy-to-use. The Trilogy 200 was designed for the home, hospital, and alternative care sites to provide invasive and noninvasive ventilator support for a wide range of pediatric and adult patients. Compared to the Trilogy 100, the Trilogy 200 has all the features of the 100, plus extra triggering sensitivity utilizing a proximal flow sensor. The extra triggering sensitivity decreases the work of breathing that provides increased comfort.

  • Easy-to-read and navigate screens.
  • Multiple prescription settings and on-screen waveforms to help optimize treatment.
  • Compact size allows for greater mobility.
  • Pressure and volume control features provide more types of therapy using just one device.
  • Use of multiple modes that gives the caregiver greater flexibility in how care can be provided.
Share:

Specifications

Dimensions
  • Height: 9.5” (23.52 cm)
  • Width: 11.2” (28.45 cm)
  • Depth: 6.6” (16.68 cm)
  • Weight: 11 lbs (5 kg)
Controls
  • Circuit types: Active Flow; Active with PAP; Passive
  • Ventilation types: Pressure and Volume
  • Volume modes: AC; SIMV (w/PS); CV; AC-MPV
  • Pressure modes: CPAP; S; S/T; T; PC-SIMV (w/PS); PC; PC-MPV; AVAPS-EA
  • IPAP: 4 to 50 cm H2O
  • PEEP/EPAP for active circuits: 0 to 25 cm H2O
  • PEEP/EPAP for passive circuits: 4 to 25 cm H2O
  • CPAP for passive circuits: 4 to 20 cm H2O
  • Pressure support differential: 0 to 40 cm H2O
General
  • Tidal volume: 0 to 2000 ml
  • Breath rate: 0 to 60 BPM for AC mode; 1 to 60 BPM for all other modes
  • Inspiratory time: 0.3 to 5.0 sec
  • Rise time: 1 to 6
  • Ramp start pressure:
    • 0 to 25 cm H2O for active circuits
    • 4 to 25 cm H2O for passive circuits
    • 4 to 19 cm H2O in CPAP
  • Ramp Length: Off; 5 to 45 min
  • Flex: Off; 1 to 3
  • Flow Trigger sensitivity: 1 to 9 l/min
  • Flow Cycle: 10 to 90 %
  • Apnea Rate: 4 to 60 BPM
  • Apnea Rate: 4 to 60 BPM
Measured Parameters
  • Respiratory Rate: 0 to 80 BPM
  • Minute Ventilation: 0 to 99 l/min
  • Vte/Vti: 0 to 2000 ml
  • Leak Rate: 0 to 2000 l/min
  • Peak Inspiratory Flow: 0 to 2000 l/min
  • Peak Inspiratory Pressure: 0 to 99 cm H2O
  • Mean Airway Pressure: 0 to 99 cm H2O
  • % Patient Triggered Breaths: 0 to 100 %
  • I:E ratio: 9.9:1 to 1:9.9